How accurate are insulin syringes at small doses? What the studies show
What 11 studies found on insulin syringe accuracy at 1 to 10 units, why the error grows at tiny volumes, and why Drawline stops at 2 units.
9 min read · By the Drawline team · Reviewed
Every fact sourced · 13 sources
Short answer: less accurate than most people think once the draw gets very small. In lab studies, people measuring 1 unit on a U-100 insulin syringe were off by about 20% or more on average. At 10 units the same syringes were close to exact. The error in the plunger and the eye stays roughly the same size, so it becomes a bigger share of a smaller draw. This review sums up what the studies measured, what they found and where their limits are.
Units here mean marks on a U-100 syringe. One unit is 0.01 mL, so 2 units is 0.02 mL (20 microliters). The studies below mostly used insulin, but the problem is the volume, so it applies to any liquid drawn into the same syringe.
The studies at a glance
| Study | Syringes | Draws tested | Main finding |
|---|---|---|---|
| Casella, Pediatrics, 1993 | 0.3 and 0.5 mL, U-100 | 0.5, 1 and 2 units | Nurses aiming for 0.5 units delivered 0.98 on average; for 1 unit, 1.64. Under 2 units the error was judged unacceptable. |
| Silva, Diabetes Care, 1996 | U-100 | 2.5 to 3.5 units | 15 caregivers overdrew by 0.22 units on average, at every dose level. |
| Gnanalingham, Arch Dis Child, 1998 | 30-unit (0.3 mL) | 1, 2, 5 and 10 units | 1 unit came out as 1.23 units (23% error). 10 units came out as 10.1. |
| Lteif, Diabetes Care, 1999 | Patients’ and parents’ syringes | Morning doses, split at 5 units | Error averaged 9.9% under 5 units and 3.2% over 5 units. |
| Keith, Clin Pediatr, 2004 | Two 30G brands, incl. half-unit marks | 1, 2 and 5 units | Syringes were “dangerously inaccurate” at 1 unit. Half-unit marks did not help. |
| Luijf, Diabetes Technol Ther, 2010 | Review of 7 studies | Low and high doses | Pens beat syringes below 5 units. Above 5 units, no clear difference. |
| Malerba, J Vet Intern Med, 2021 | 0.3 mL U-100 and U-40 | 0.5 to 16 units | At 0.5 units the 0.3 mL syringe varied by 12% from draw to draw. At 4 units the average was on target. |
| Kondo, Diabetol Int, 2024 | Syringe and vial | 2, 6 and 10 units | 33 nurses missed a 2-unit target by 0.6 units (median), up to 2.3 units. |
| Henning, JAVMA, 2026 | 0.5 mL insulin, 1 mL | 0.02, 0.12 and 0.43 mL | Errors of 20% or more were common, worst at the smallest volume and with larger needles. |
| Geus, Int J Retina Vitreous, 2026 | Six syringes, incl. a U-100 insulin syringe | 50 µL (5 units) | Only 45% of draws by 23 trained staff landed within 10% of the target. |
| Arnott, Med J Aust, 1982 | U-100 with and without dead space | Low doses, two insulins mixed | Syringes with dead space gave errors over 100% at low doses. |
What each study measured
Casella and colleagues, 1993 (Johns Hopkins). The team weighed doses drawn by pediatric nurses and by parents of children with diabetes. Nurses aiming for 0.5, 1 and 2 units delivered 0.98, 1.64 and 2.15 units on average. A 0.3 mL syringe did no better than a 0.5 mL one. Parents were closer as a group at 1 unit, but only 3 of 7 could deliver it with acceptable accuracy and precision. The authors concluded that draws under 2 units have an unacceptably large error and advised diluted insulin in hospital when a dose is under 2 units. Limits: a small group, insulin only, and the full methods sit behind a paywall, so this summary uses the abstract.
Silva and colleagues, 1996. 15 caregivers drew repeated doses between 2.5 and 3.5 units, some only 0.25 units apart. Every dose leaned high, by 0.22 units on average. The caregivers were consistent with themselves: they could tell a 0.25-unit change. When two people shared the job, the combined spread was about 40% larger than one person’s. Limits: a small sample and one setting.
Gnanalingham and colleagues, 1998. Five nurses drew 1, 2, 5 and 10 units with 30-unit syringes, 15 times each, against two insulin pens. Each dose was weighed. At 1 unit the syringe delivered 1.23 units on average, a 23% error. At 10 units it delivered 10.1. The authors called all three devices, especially syringes, unacceptably inaccurate at 1 unit. Limits: five syringes and five nurses, one insulin type.
Lteif and Schwenk, 1999 (Mayo Clinic). 32 children with diabetes and the parents of 16 more measured their usual morning doses three times, using a traced saline solution in place of insulin. For doses under 5 units, the syringe error averaged 9.9%, against 4.9% for pens. Over 5 units the two were close (3.2% and 2.2%). Limits: the study grouped doses above and below 5 units rather than testing set volumes.
Keith and colleagues, 2004 (Cleveland Clinic). One investigator drew 1, 2 and 5 units 15 times each with two brands of syringe, three pens and a pump, and a second weighed them. Syringes were less accurate than pens and the pump at 1 and 2 units, and the authors called them dangerously inaccurate at 1 unit. Syringes and pens with half-unit marks did not improve accuracy or precision. Limits: one person drawing, so it shows the device more than the range of real users.
Luijf and DeVries, 2010. This review pooled 7 studies comparing pens with syringes. They agreed that pens are more accurate below 5 units, and found no clear benefit above 5 units. Limits: a narrative review of small studies, most of them in children.
Malerba and colleagues, 2021 (veterinary). Five veterinarians drew 0.5 to 16 units 25 times each with 0.3 mL U-100 and 1 mL U-40 syringes, against six pens. At 0.5 units, the 0.3 mL syringe varied by 12% from one draw to the next. By 4 units its average was not significantly different from the target. Limits: experienced users and a lab bench, not home use.
Kondo and colleagues, 2024 (Japan). 33 ward nurses made 198 draws from vials at targets of 2, 6 and 10 units. The median miss was 0.6 units at 2 and 6 units and 0.7 units at 10 units. The worst miss at the 2-unit target was 2.3 units, more than the dose itself. Errors were larger when nurses skipped the air-bubble check. Experience and time of day made no difference. Limits: one hospital, and the abstract does not give the syringe size.
Henning and colleagues, 2026 (veterinary). 30 people with three levels of experience filled 180 syringes with water at 0.02, 0.12 and 0.43 mL. Errors of 20% or more were common, most of all at the smallest volume and with larger syringe-and-needle sets. The smallest needle gave the tightest results at the smallest volume. Beginners were about twice as likely as experts to land within 20%. Limits: water, not a drug, and a clinic setting.
Geus and colleagues, 2026 (eye injections). 23 specialist staff drew a 50 µL dose, the same as 5 units on a U-100 syringe, with six syringe types. Draws ran high (median 55.1 µL), and only 45.4% landed within 45 to 55 µL. About a fifth of the spread came from differences between people. No syringe type was clearly more accurate than the rest. Limits: most syringes tested were 1 mL syringes, not insulin syringes.
Arnott and colleagues, 1982. People with diabetes drew two insulins into one syringe. Syringes with dead space (liquid left in the hub and needle) gave errors over 100% at low doses, and dose-to-dose results were poor. Syringes with minimal dead space fixed this. Limits: an older study; most insulin syringes now have the needle fixed to the barrel.
Manufacturing tolerance versus drawing error
The syringe itself is rarely the main problem. Insulin syringes are made and tested to the international standard ISO 8537. BD’s 2019 FDA clearance for its 0.3, 0.5 and 1 mL insulin syringes lists volumetric accuracy and dead space testing under that standard, and all three sizes have permanently attached needles. The same filing lists the scales: 1-unit and half-unit marks on the 0.3 mL syringe, 1-unit marks on the 0.5 mL and 2-unit marks on the 1 mL. We did not read the paid text of ISO 8537, so this page does not quote its tolerance figures.
In the studies above, most of the error came from people: reading the line, air bubbles and how far the plunger moved. The best evidence for this is that trained nurses and specialist staff made errors as large as parents did.
Reports of real errors
FDA’s July 2024 alert on compounded semaglutide described patients drawing 50 units when told to draw 5 (0.05 mL). It also said some patients received syringes “significantly larger than the prescribed volume,” and asked providers and compounders to supply the right syringe size for the dose. Read more in the 5 units vs 50 units mistake.
What this means for measuring
- Error is roughly a fixed amount, so it grows as a share. In Silva’s study people overdrew by about 0.2 units at every level. That is 2% of a 10-unit draw, 10% of a 2-unit draw and 20% of a 1-unit draw.
- A smaller syringe helps you read the line, but does not fix tiny volumes. A 0.3 mL syringe spaces the marks further apart. Even so, Casella found 0.3 mL no better than 0.5 mL under 2 units, and Keith found half-unit marks did not help.
- Read the line the same way every time. Hold the syringe at eye level and read at the flat top edge of the plunger seal. See how to read syringe units.
- Check for air bubbles. In Kondo’s study, skipping this check was the one habit linked to bigger errors.
- Match the syringe to the scale. A U-40 syringe holds 2.5 times more per line than a U-100. See U-40 vs U-100 syringes.
- Ask your pharmacist which syringe fits your prescription, and how to measure it, before the first draw from a new vial.
Why Drawline won’t calculate a draw under 2 units
Every Drawline calculator shows Do not draw, with no number and an empty syringe, when the answer is under 2 units (0.02 mL). Here is how that rule lines up with the evidence.
- The evidence supports 2 units as a floor. Casella’s team set the same line: they found draws under 2 units had unacceptable error and advised dilution below it. Gnanalingham, Keith and Malerba all found syringes worst at 1 unit and below.
- 2 units is not a guarantee. Errors do not stop at 2 units. Kondo’s nurses missed a 2-unit target by 0.6 units at the median, about 30%. Lteif found about 10% error under 5 units, and the 2010 review found pens beat syringes up to 5 units. If anything, the evidence would support a higher line for accuracy. We chose 2 units as the point below which we will not show a number at all, and the studies suggest that draws between 2 and 5 units still deserve extra care.
- A tiny answer often means a typo. The most common reason for a draw under 2 units in a calculator is a unit mix-up, such as mcg typed as mg. Refusing to show a number makes you stop and check.
When you mix a powder vial, more bacteriostatic water spreads the same amount over more volume, so the same dose sits on a bigger number. The BAC water calculator shows the arithmetic. For a liquid vial a pharmacy dispensed, ask your pharmacist how to measure it. Diluting a prescription product is a pharmacist’s or prescriber’s decision.
None of this says what dose anyone should take. Drawline only does the arithmetic for the dose your prescriber wrote. The mg to units converter shows the volume and units for that dose, and when the answer is under 2 units it tells you to stop and check.
This review is for research and education. It is not medical advice. Questions about your own prescription belong with your prescriber or pharmacist.
Sources: Casella SJ et al. Accuracy and precision of low-dose insulin administration. Pediatrics 1993;91(6):1155-7 (abstract); Silva SR et al. Can caretakers of children with IDDM accurately measure small insulin doses and dose changes? Diabetes Care 1996;19(1):56-9 (abstract); Gnanalingham MG et al. Accuracy and reproducibility of low dose insulin administration using pen-injectors and syringes. Arch Dis Child 1998;79(1):59-62; Lteif AN, Schwenk WF. Accuracy of pen injectors versus insulin syringes in children with type 1 diabetes. Diabetes Care 1999;22(1):137-40 (abstract); Keith K et al. Accuracy and precision of low-dose insulin administration using syringes, pen injectors, and a pump. Clin Pediatr 2004;43(1):69-74 (abstract); Luijf YM, DeVries JH. Dosing accuracy of insulin pens versus conventional syringes and vials. Diabetes Technol Ther 2010;12 Suppl 1:S73-7 (abstract); Malerba E et al. The accuracy and precision of insulin administration using human and veterinary pen-injectors and syringes. J Vet Intern Med 2021;35(3):1255-64 (abstract); Kondo M et al. Analysis of factors affecting the accuracy of low-dose insulin dosage using syringes and vials. Diabetol Int 2024;15(3):544-9 (abstract); Henning J et al. User experience and syringe-and-needle combination impacts accuracy and precision of small-volume medications. JAVMA 2026;264(5):599-606 (abstract); Geus PF et al. Performance characterization of syringes used for intravitreal injections. Int J Retina Vitreous 2026 (abstract); Arnott RD et al. Insulin syringes: dangers of dead space. Med J Aust 1982;2(1):39-40 (abstract); FDA 510(k) K190054: BD Insulin Syringe, 0.3, 0.5 and 1 mL (October 21, 2019); FDA: dosing errors associated with compounded injectable semaglutide products (July 26, 2024).
This guide is for research and educational use. It is not medical advice. Confirm anything that affects your health with a licensed professional.
Last updated · Revision history
First published . No content changes since. Fact corrections are listed in the corrections log, and every changed page on What changed.
Spot a mistake or an out-of-date fact? Report an error. We check every report against the sources.